Dry-Wet Integrated Bridge Pendant / Surgical Tower
1. Working power supply: AC 220 V, 50 Hz.
2. Beam length: 2500–3000 mm (actual dimensions subject to on-site measurement by the user); 1 lighting lamp.
3. Optional brake device: the equipment has no drift when braked, and can be moved easily when released.
4. Surgical tower: 1 unit (left-right movement distance 300 mm, adjustable). Configuration as follows:
(1) Instrument platforms: 2 layers (height adjustable), 660 mm × 450 mm, with silicone gel corner protectors and anti-collision design;
(2) Standard gas terminal configuration: (2 oxygen, 2 vacuum, 2 medical air), configurable according to hospital requirements;
a. Different colors and shapes of connectors to prevent misconnection;
b. More than 20,000 insertion and withdrawal cycles;
c. Secondary sealing with three states (on, off, unplugged), and can be serviced while gas is supplied;
(3) Multi-function power sockets: ≥10 pcs (configurable according to hospital), 220 V;
(4) Grounding terminals: 2 pcs;
(5) Network interface: 1 pc;
(6) Drawer: 1 pc;
(7) Stainless steel IV pole: 1 pc;
(8) Rotation: 0–340°;
(9) Net load capacity of the box: ≥200 kg.